Function | |||
Interactive Interface | LCD, Keypad | ||
Measurement Range | SBP:60~260mmHg, DBP:30~195mmHg; Pulse Rate: 30~200bpm | ||
Accuracy |
Pressure:±3mmHg Pulse Rate:±5% |
||
Reproducibility | Pressure:≤3mmHg | ||
Method of Measurement | Oscillometry with linear deflation | ||
Record Capacity |
ABP-010 & ABP-020 :200 records; ABP-011 & ABP-021 :400 records; |
||
Physical Specification | |||
Monitor Dimension | 113 x 75 x 26 mm | ||
Monitor Weight | 168g without battery | ||
Monitor Ingress Protection | IP22 | ||
Cuff Size |
M5402BM: arm circumference 20~24cm M5403BM: arm circumference 24~32cm M5404BM: arm circumference 32~38cm |
||
Collision | Still works in case fall on hard surface from 50mm height in any axial direction | ||
Operating Environment | |||
Temperature | 5°C ~ +40°C | ||
Humidity | 10% ~ 95% RH(no condensed) | ||
Pressure | 800hPa ~ 1060hPa | ||
Storage and Transport Environment | |||
Temperature | -25°C ~ +70°C | ||
Humidity | ≤95% RH(no condensed) | ||
Pressure | 560hPa ~ 1060hPa | ||
Battery | |||
Type | 2 AA(LR6) alkaline or high capacity Ni-HM batteries | ||
Supply Voltage | 2.4~3.2V | ||
Measurement Capacity | 200~400 measurements per battery |